Personnel moved rapidly in the vicinity of open sterile units and instruments, which disrupted the airflow and increased the risk of bringing lesser quality air into the ISO 5 classified aseptic processing area.
Inspection Record
Johnson Memorial Cancer Center — FDA 483 Findings
Inspectors Erik W Koester · John P MistlerThe inspectors who signed the published document.Inspector lookup
View the regulator's official source The source document is always the basis for judgement.
Findings
You did not make adequate product evaluation and take remedial action where actionable microbial contamination was found to be present in an area adjacent to the ISO 5 classified aseptic processing area during aseptic production.
You did not make adequate product evaluation and take remedial action where actionable microbial contamination was found to be present in the ISO 5 classified aseptic processing area during aseptic production.
Your facility design allowed the influx of poor quality air into a higher classified area.
Disinfecting agents and cleaning wipes used in the ISO 5 classified aseptic processing areas were not sterile.
You had inadequate HEP A filter coverage and airflow over the area to which sterile product was exposed.