Critical process parameters for FFS should be determined during equipment qualification and should include, but are not limited to:

i. Settings for uniform package dimensions and cutting in accordance with validated parameters.

ii. Setting, maintenance and monitoring of validated forming temperatures (including pre-heating and cooling), forming times and pressures as relevant.

iii. Setting, maintenance and monitoring of validated sealing temperatures, sealing temperature uniformity across the seal, sealing times and pressures as relevant.

iv. Environmental and product temperature.

v. Batch-specific testing of package seal strength and uniformity.

vi. Settings for correct filling volumes, speeds and uniformity.

vii. Settings for any additional printing (batch coding), embossing or debossing to ensure that unit integrity is not compromised.

viii. Methods and parameters for integrity testing of filled containers (see paragraph 8.22).

Source: European Commission, DG SANTE · © European Union. Reuse is authorised provided the source is acknowledged (Commission Decision 2011/833/EU on the reuse of Commission documents). · retrieved 2026-11-15

Weekly Newsletter

Get new findings like these every Monday

Every Monday we email the week's new findings and actions from the FDA, EMA, Health Canada and MFDS. Each one links to its record on our site, where you'll find the regulator's original document. We'll also let you know when the site gains new features. Unsubscribe any time with one click.

See recent findings first

Sent every Monday · a confirmation email follows when you subscribe · unsubscribe from the footer of any email.