EU GMP Annex 1 8.52
8. Production and Specific Technologies › Sterilisation by heat
Section text
The whole of the load should reach the required temperature before measurement of the sterilising time-period starts. For sterilisation cycles controlled by using a reference probe within the load, specific consideration should be given to ensuring the load probe temperature is controlled within defined temperature range prior to cycle commencement.