Skip to content

EU · EudraLex Vol.4

EU GMP Part I, Chapter 6 6.17

Testing

Section text

The tests performed should be recorded and the records should include at least the following data:

i. Name of the material or product and, where applicable, dosage form;

ii. Batch number and, where appropriate, the manufacturer and/or supplier;

iii. References to the relevant specifications and testing procedures;

iv. Test results, including observations and calculations, and reference to any certificates of analysis;

v. Dates of testing;

vi. Initials of the persons who performed the testing;

vii. Initials of the persons who verified the testing and the calculations, where appropriate;

viii. A clear statement of approval or rejection (or other status decision) and the dated signature of the designated responsible person;

ix. Reference to the equipment used.

Source: European Commission, DG SANTE · © European Union. Reuse is authorised provided the source is acknowledged (Commission Decision 2011/833/EU on the reuse of Commission documents). · retrieved 2026-11-15