The use of sporicidal agents in the ISO 5 classified aseptic processing area was inadequate.
Inspection Record
PharmScript, LLC — FDA 483 Findings
View the regulator's official source The source document is always the basis for judgement.
Findings
Personnel touched equipment or other surfaces located outside of the ISO 5 classified aseptic processing area with gloved hands and then engaged in aseptic processing without changing or sanitizing gloves.
Personnel engaged in aseptic processing were obse1ved with exposed hands, exposed wrists and exposed hair.
You produced beta-lactam drugs without providing adequate containment, segregation and cleaning of work surfaces to prevent cross-contamination.