Personnel were observed moving quickly in a critical area or in an area immediately adjacent to a critical area likely causing disruption of unidirectional airflow.
Inspection Record
Richmond VA Medical Center — FDA 483 Findings
Inspectors Tekalign Wondimu · Michael O IdowuThe inspectors who signed the published document.Inspector lookup
View the regulator's official source The source document is always the basis for judgement.
Findings
Personnel performed aseptic manipulations with exposed hair or skin.
Lack of disinfection of equipment at each transition from areas of lower quality air to areas of higher quality air.
Failure to appropriately and regularly clean and disinfect or sterilize equipment located in the ISO 5 area.
Smoke studies were not performed under dynamic conditions.
Personnel were observed conducting aseptic manipulations where the movement of "first air" in the ISO 5 area is blocked or disrupted.
Use of a sporicidal agent in the facility's ISO 5 areas was improper.
Use of a disinfectant in a manner insufficient to achieve adequate levels of disinfection.