Your firm failed to perform required tests on bulk medical(b) (4) USP received for production.
Inspection Record
Conroe Welding Supply, Inc — FDA 483 Findings
Inspectors Timothy H Vo · Rodney L TinzieThe inspectors who signed the published document.Inspector lookup
View the regulator's official source The source document is always the basis for judgement.
Findings
There is no quality unit.
Batch production records do not include documentation that each significant step in the packaging of the medical gas was performed.
The quality unit failed to review and approve manufacturing production and control records, including those for packaging and labeling, to determine compliance with all established, approved written procedures before batches were released or distributed.
Your firm failed to establish a written program designed to ensure proper performance of automatic, mechanical, and electronic equipment used to produce medical (b) (4) Specifically, The new (b) (4) equipment (Db) (4) Pumping System) used for production and the (DB) (4) used for testing lack a written program to ensure proper performance and maintenance.