Failure to conduct media fills that closely simulate aseptic production operations under the worst-case, most-challenging, and stressful conditions.
Inspection Record
Victory Medical Center Pharmacy — FDA 483 Findings
View the regulator's official source The source document is always the basis for judgement.
Findings
The <Redacted B4> intended to render final product sterile is not adequate and not appropriate for pharmaceutical use.
Sterile drugs and Materials were exposed to lower than ISO 5 quality air.
The facility design was observed to allow the influx of lesser quality air into a classified area containing higher quality air.
Use of ingredients not intended for pharmaceutical use in sterile drug production.
Failure to conduct post-use <Redacted B4> testing on <Redacted B4> used to sterilize drug products.
Lack of routine environmental monitoring in the ISO 5 area and classified areas.
Pr o d uce r o f St e r i l e a n d Non- s t e r i l e Dr u g Pr o d ucts Failure to appropriately and regularly clean and disinfect or sterilize equipment located in the ISO 5 area.
Production areas have difficult to clean or contain porous, particle generating, or visibly dirty equipment or surfaces.
Personnel were observed manually contacting the inner surface of a sterile container or closure.
Smoke studies were not performed under dynamic conditions.
Similar findings at other companies
GRM selected findings from the same type of public document whose wording is close to a finding in this document. One per company.
- Strive Pharmacy LLC2026-09-28 · US FDASimilar to finding 11 Smoke studies were inadequately performed under dynamic conditions.
- XeCare LLC2026-06-17 · US FDASimilar to finding 5 Use of ingredients not intended for pharmaceutical use in non-sterile drug production.
- Greenwich Rx2026-09-09 · US FDASimilar to finding 4 The facility design was observed to allow the influx of lesser quality air into a classified area containing higher quality air.
- Richmond VA Medical Center2026-05-12 · US FDASimilar to finding 11 Smoke studies were not performed under dynamic conditions.
A record as of the US FDA document published on 2026-10-21. Check the agency's announcements for any later corrective action, and rely on the original text. Translation and classification are machine-made.