Findings

1Other quality system

The acceptance status ofproduct was not clearly identified throughout installation and servicing ofthe product.

2Other quality system

Not all ofthe actions needed to correct and prevent the recurrence ofnonconforming product and other quality problems have been identified.

3Other quality system

A process whose results cannot be fully verified by subsequent inspection and test has not beenadequately validated and approved according to established procedures.

4Deviation, CAPA, and investigation

A validated process was not revalidated when changes or process deviations occurred.

5Other quality system

Procedures that describe the review and disposition process for nonconforming product were not implemented.

6Other quality system

The organization structureis not adequateto assure that quality system requirements are fully met.

About this record

Extracted automatically from the document US FDA published on 2024-01-17. The source is available at the link above.

This is a record of that moment. Inspection findings are usually followed by the company's response and corrective action, but this page does not know how that turned out. Do not read it as the current state; check the regulator's official announcements for the latest status.

Translation and classification are automated and may differ in nuance from the source.

Boston Scientifc Corporation company profile — full history US FDA documents, 2024 Go to Findings search