Reference standards used by the firm for analysis are accepted and used as 100% pure, although certificates of analysis, in certain instances, do not contain adequate data on the purity or potency of the standard.
Inspection Record
Forest/Inwood Laboratories, Inc. — FDA 483 Findings
View the regulator's official source The source document is always the basis for judgement.
Findings
The following was noted regarding the multiprobe robotic micro-pipettor, and the automated solid phase extractors: a) There are no written SOPs for the calibration of either the multiprobe robotic micro-pippettor, or the automated solid phase extractor units.
The stock solutions for Sgiiigemeiidilillke made on 8/22/95, and for ier made on 3/4/96, were each used for the preparation of both QC and standard curve solutions.
The reported assay precision in the analytical runs was not accurate because the .
The firm’s acceptance criteria for the injection of the mix solution at the beginning of the analytical run is not adequate to assure reproducibility of the chromatographic system.
For para-hydroxyrimantadine hydrochloride used in study FLU-PK-03-000, the calculated percentages of axial and equatorial isomers were 47.4 and 52.6 respectively.
Firm's SOPs allow for non-USP analytical reference standards, which are not covered by a supplier's assigned expiration date, to be used for up to giiiiaeenaiiivess without re-testing or other available supporting stability data.
Reference standards are weighed on an analytical balance and the weights recorded to 4 decimal places (eg.
Reference standards are used for analysis with an assumed 100% potency regardless of the actual potency/purity.
There was no established criteria fordetermining assay interference.
Firm's SOPs require calibration of the manually operated digital micro-pipettes
Freezer and refrigeration units used to store study specimens and test solutions are not monitored by any continuous time/temperature recording devices.
Reference standards used for analysis are not dried prior to weighing nor are loss on drying or water factors used to correct the weights.
Firm’s SOPs allow for the stability of stock solutions to be checked by chromatographic comparison with the mix solution.