A validated process was not performed by a qualified individual.
Inspection Record
Stryker Instruments — FDA 483 Findings
Inspectors Charles M. Spyr · Martha Sullivan Myrick · Kimberl Lewandowskiwalker and 1 moreThe inspectors who signed the published document.Inspector lookup
View the regulator's official source The source document is always the basis for judgement.
Findings
Rework and reevaluation activities have not been fully documented inthe device history record.
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Procedures were not established for reviewing sampling methods for adequacy for their intended use.
An MDRreport was not submitted w i t h 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious ijury if the malfunction were to recur.
Potential suppliers and contractors were not sufficiently evaluated and selected on the basis of their ability to meet specified requirements.
Complaint handling procedures for receiving complaints have not been implemented.
Procedures for addressing the identification, evaluation, and investigation of nonconforming product were not defined.
Similar findings at other companies
GRM selected findings from the same type of public document whose wording is close to a finding in this document. One per company.
- Invacare Corporation2015-07-11 · US FDASimilar to finding 5 An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has…
- H & P Industries, Inc.2015-07-11 · US FDASimilar to finding 3 A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- Diagast2024-04-19 · US FDASimilar to finding 5 An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has…
- Legacy Pharmaceutical Packaging LLC2020-01-16 · US FDASimilar to finding 2 Rework and reevaluation activities have not been documented in the device history record.
- Unico Holdings, Inc.2015-07-11 · US FDASimilar to finding 3 A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established…
A record as of the US FDA document published on 2024-01-17. Check the agency's announcements for any later corrective action, and rely on the original text. Translation and classification are machine-made.