The calibration, inspection, and/or qualification of the equipment, including computerized systems, was inadequate.
Inspection Record
Serum Institute of India Pvt. Ltd. — Health Canada Inspection Findings
View the regulator's official source The source document is always the basis for judgement.
Findings
The operators'aseptic practices were inadequate to maintain sterility.
Investigations into deviations, reports, and/or follow-up actions were inadequate.
The company did not verify the sterilization process at scheduled intervals or after significant modifications.
Evaluation of production processes, equipment, and/or materials was inadequate. The samples taken for sterility testing were not representative of the whole batch.
The company did not adequately validate radiation sterilization.
The company's monitoring and documentation of moist heat sterilization cycles was inadequate.