FDA looks less at whether a water system is installed than at whether validation and monitoring confirm that it consistently produces water suitable for its use. The findings below are quoted from warning letters and FDA 483s.
What inspectors cited
1The water system was not shown to be validated
You failed to adequately validate your water production process to ensure the system was consistently producing water of appropriate quality for its intended use.
FDA Warning Letter · Naturich Cosmetique Labs · 2025-10-07 · View document
You were unable to provide adequate documentation demonstrating that your water system is validated.
FDA Warning Letter · Nature’s Fusions LLC · 2025-07-22 · View document
2Water was not checked for the absence of objectionable microorganisms
For example, your non-sterile semaglutide and tirzepatide API are intended for sterile injectable drug products, and your (b)(4) water system has not been evaluated for the absence of objectionable microorganisms.
FDA Warning Letter · Harbin Jixianglong Biotech Co., Ltd. · 2026-05-19 · View document
Specifically, The firm's non-sterile topical drug products and [redacted] Water, used in producing the non-sterile drug products, are not routinely tested for objectionable microorganisms such as Burkholderia cepacia complex (BCC).
FDA 483 · Sincerus Florida, LLC · 2026-07-22 · View document
3The system had dead legs and water that did not circulate
For example, your water system included a dead leg, was not continuously circulating, and was not subjected to validated sanitization processes.
FDA Warning Letter · Evi Labs LLC · 2025-06-03 · View document
For example, your non-circulating (b)(4) water system included dead legs, which could foster the development of biofilms. When not in use, water from this system sat stagnant, including in a line that was used to dispense water for your drug product.
FDA Warning Letter · Multaler et Cie SAS · 2024-11-12 · View document
4Total organic carbon testing was not performed
You lacked water system validation and failed to conduct total organic carbon (TOC) testing to ensure that your component water met (b)(4) Water United States Pharmacopeia (USP) monograph specifications.
FDA Warning Letter · Medical Chemical Corporation · 2025-07-22 · View document
You lacked conductivity and total organic carbon limits and testing and did not test for specific objectionable microorganisms.
FDA Warning Letter · Nature’s Fusions LLC · 2025-07-22 · View document
5Adequate monitoring of the water system was not demonstrated
Based on the records and information you provided, you failed to demonstrate that your (b)(4) water system is adequately monitored to ensure that it consistently produces water suitable for use in the manufacture of APIs intended for further processing into sterile drug products.
FDA Warning Letter · Hangzhou Yiqi Biotechnology Co., Ltd · 2026-05-19 · View document
Furthermore, you failed to demonstrate that your water system was adequately monitored to ensure it consistently produced water that met appropriate chemical and microbial quality standards.
FDA Warning Letter · Naturich Cosmetique Labs · 2025-10-07 · View document
Questions to check against your procedures
- Can you present validation documents showing that the water system consistently produces water of suitable quality for its use?
- Is water routinely tested for the absence of objectionable microorganisms such as Burkholderia cepacia complex?
- Has water used to manufacture APIs intended for sterile injectables been evaluated for objectionable microorganisms?
- Is the water system free of dead legs and of lines where water sits stagnant when not in use?
- Does the water system circulate continuously, and is it subjected to a validated sanitization process?
- Are limits set and tests run for total organic carbon and conductivity, so that monitoring shows the water meets chemical and microbial quality standards?
Compiled by AI from public findings and checked by a person. Quotations are verbatim from the original documents; (b)(4) marks text redacted in the original. As of 2026-11-15.