EU GMP Annex 1 8.77
8. Production and Specific Technologies › Sterilisation with ethylene oxide
Section text
Critical process parameters that could be considered as part of the sterilisation process validation and routine monitoring include, but are not limited to:
i. EO gas concentration.
ii. Pressure.
iii. Amount of EO gas used.
iv. Relative humidity.
v. Temperature.
vi. Exposure time.