EU · EudraLex Vol.4
EU GMP Annex 1 8.77
8. Production and Specific Technologies › Sterilisation with ethylene oxide
조항 원문
Critical process parameters that could be considered as part of the sterilisation process validation and routine monitoring include, but are not limited to:
i. EO gas concentration.
ii. Pressure.
iii. Amount of EO gas used.
iv. Relative humidity.
v. Temperature.
vi. Exposure time.
본문은 원문(영어)입니다. 국문 번역은 준비 중입니다.
출처: European Commission, DG SANTE · © European Union. Reuse is authorised provided the source is acknowledged (Commission Decision 2011/833/EU on the reuse of Commission documents). · 수집일 2026-11-15