Your firm failed to routinely calibrate, inspect, or check according to a written program designed to assure proper performance of automatic, mechanical, electronic equipment, or other types of equipment, including computers, used in the manufacture, processing, packing, and holding of a drug product (21 CFR 211.68(a)). You failed to adequately inspect and maintain your AHUs to ensure air filters would be effective at preventing contamination. The procedure for HEPA filter…
Inspection Record
Granules India Limited — FDA Warning Letter Findings
View the regulator's official source The source document is always the basis for judgement.
Findings
Your firm failed to establish and follow adequate written procedures for cleaning and maintenance of equipment (21 CFR 211.67(b)). Our investigators observed significant contamination in multiple (b)(4) ducts of non-dedicated (b)(4) used in the preparation of (b)(4) for finished drug products manufactured at your facility. While filters were installed to prevent contamination, inadequate cleaning and maintenance processes rendered them ineffective. Swab samples collected…
Your firm failed to maintain buildings used in the manufacture, processing, packing, or holding of drug products in a good state of repair (21 CFR 211.58). Bird droppings and feathers were observed during the inspection in the AHU area, specifically on the air purification units, (b)(4) ducts, a (b)(4) tank, a cleaning (b)(4) , and on the floors surrounding multiple (b)(4) inside your drug manufacturing facility. The conditions in your AHU area raise concerns about potential…
About this record
Extracted automatically from the document US FDA published on 2025-03-04. The source is available at the link above.
This is a record of that moment. Inspection findings are usually followed by the company's response and corrective action, but this page does not know how that turned out. Do not read it as the current state; check the regulator's official announcements for the latest status.
Translation and classification are automated and may differ in nuance from the source.