Findings

1Validation and qualification

ISO 5 Areas are not certified under dynamic conditions.

2Equipment and facility

Your facility design allowed the influx of poor quality air into a higher classified area.

3Aseptic processing and sterility assurance

The use of sporicidal agents in the cleamooms and ISO 5 classified aseptic processing area was inadequate.

4Equipment and facility

The ISO 7 classified areas have difficult to clean, particle -generating equipment or surfaces.

5Aseptic processing and sterility assurance

You did not make adequate product evaluation and take remedial action where actionable microbial contamination was found to be present in an area adjacent to the ISO 5 classified aseptic processing area during aseptic production.

6Equipment and facility

Your facility was designed and/or operated in a way that permits poor flow of personnel.

7Aseptic processing and sterility assurance

Disinfecting agents and cleaning pads used in the ISO 5 classified aseptic processing areas were not sterile.

About this record

Extracted automatically from the document US FDA published on 2019-11-05. The source is available at the link above.

This is a record of that moment. Inspection findings are usually followed by the company's response and corrective action, but this page does not know how that turned out. Do not read it as the current state; check the regulator's official announcements for the latest status.

Translation and classification are automated and may differ in nuance from the source.

Roger Williams Medical Center company profile — full history US FDA documents, 2019 Go to Findings search