Inspection Record

RC Outsourcing LLC — FDA 483 Findings

US FDAInspected 2024-09-27Published 2024-12-12 8 findingsComplaint and recall handlingAseptic processing and sterility assuranceProcess validationOther quality systemLaboratory and QC controlsMaterial and supplier control

Inspectors Taichun QinThe inspectors who signed the published document.Inspector lookup

View the regulator's official source The source document is always the basis for judgement.

Findings

1Complaint and recall handling

Complaint procedures are deficient in that they do not include provisions that allow for the review and determination of an investigation by the quality control unit.

2Aseptic processing and sterility assurance

Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions.

3Process validation

Written production and process control procedures are not followed in the execution of production and process control functions.

4Other quality system

Reserve drug product samples are not appropriately identified and representative of each lot or batch of drug product.

5Laboratory and QC controls

Laboratory controls do not include the establishment of scientifically sound and appropriate specifications designed to assure that conform to appropriate standards of identity, strength, quality and purity.

6Laboratory and QC controls

Laboratory records do not include the initials or signature of a second person showing that the original records have been reviewed for accuracy, completeness and compliance with established standards.

7Aseptic processing and sterility assurance

Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written and followed.

8Material and supplier control

There is a failure to thoroughly review the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.

About this record

Extracted automatically from the document US FDA published on 2024-12-12. The source is available at the link above.

This is a record of that moment. Inspection findings are usually followed by the company's response and corrective action, but this page does not know how that turned out. Do not read it as the current state; check the regulator's official announcements for the latest status.

Translation and classification are automated and may differ in nuance from the source.

Other records for this company

RC Outsourcing LLC company profile — full history US FDA documents, 2024 Go to Findings search