The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Inspection Record
Collaborative Care Diagnostics, LLC — FDA 483 Findings
Inspectors Tyler R CourtneyThe inspectors who signed the published document.Inspector lookup
View the regulator's official source The source document is always the basis for judgement.
Findings
One or more processes for risk management in product realization have not been adequately documented.
The medical device(s) label did not have a unique device identifier (UDI) present in easily readable plain-text and automatic identification and data capture (AIDC) technology.
Review of design and development changes did not include evaluation of the effect of the changes on constituent parts and product in process or already delivered, inputs or outputs of risk management and product realization processes.
Similar findings at other companies
GRM selected findings from the same type of public document whose wording is close to a finding in this document. One per company.
- Cardionxt2025-12-15 · US FDASimilar to finding 1 The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device…
- Snap Diagnostics, LLC.2026-02-13 · US FDASimilar to finding 2 One or more processes for risk management in product realization have not been adequately documented.
A record as of the US FDA document published on 2026-10-20. Check the agency's announcements for any later corrective action, and rely on the original text. Translation and classification are machine-made.