Risk analysis is inadequate.
Inspection Record
Cardionxt — FDA 483 Findings
Inspectors Hammam A AbdallahThe inspectors who signed the published document.Inspector lookup
View the regulator's official source The source document is always the basis for judgement.
Findings
Procedures for corrective and preventive action have not been adequately established.
Procedures have not been adequately established to control product that does not conform to specified requirements.
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Similar findings at other companies
GRM selected findings from the same type of public document whose wording is close to a finding in this document. One per company.
- Implantica CE Reflux Limited2025-10-24 · US FDASimilar to finding 2 Procedures for corrective and preventive action have not been adequately established.
- American Contract Systems, Inc2018-05-24 · US FDASimilar to finding 3 Procedures have not been adequately established to control product that does not conform to specified requirements.
- Invacare Corporation2015-07-11 · US FDASimilar to finding 2 Procedures for corrective and preventive action have not been adequately established.
- Collaborative Care Diagnostics, LLC2026-02-09 · US FDASimilar to finding 6 The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device…
- Bausch and Lomb, Inc2015-07-24 · US FDASimilar to finding 4 Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary .
A record as of the US FDA document published on 2026-11-06. Check the agency's announcements for any later corrective action, and rely on the original text. Translation and classification are machine-made.