The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.
Inspection Record
Diagast — FDA 483 Findings
Inspectors Creighton T Tuzon · Richard T Riggie · Elias N Paz AlonzoThe inspectors who signed the published document.Inspector lookup
View the regulator's official source The source document is always the basis for judgement.
Findings
Procedures for the control of storage areas and stock rooms have not been adequately established.
Failure to submit a biological deviation report.
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
In-process inspections, tests, or other verification activities and approvals were not documented.
Risk analysis is inadequate and is incomplete.