Inspection Record

Isologic Innovative Radiopharmaceuticals Ltd. — Health Canada Inspection Findings

Health CanadaInspected 2025-07-15 6 findingsDeviation, CAPA, and investigationComputer system validationEquipment and facilityOther quality systemContamination control

View the regulator's official source The source document is always the basis for judgement.

Findings

1Deviation, CAPA, and investigation

Investigations into deviations, reports, and/or follow-up actions were inadequate.

2Computer system validation

The calibration, inspection, and/or qualification of the equipment, including computerized systems, was inadequate.

3Equipment and facility

The company did not design or maintain the premises and equipment to minimize microbial contamination and the number of airborne particles in clean areas. The sanitation programs were not adequately designed to identify and/or reduce the risks of contamination.

4Other quality system

The controls were inadequate for creating, modifying, reviewing, storing, and/or retrieving records.

5Other quality system

The environmental sampling methods were inadequate.

6Contamination control

The company did not adequately demonstrate that air-flow patterns did not pose a risk of contamination.

About this record

Extracted automatically from the document of the Health Canada inspection on 2025-07-15. The source is available at the link above.

This is a record of that moment. Inspection findings are usually followed by the company's response and corrective action, but this page does not know how that turned out. Do not read it as the current state; check the regulator's official announcements for the latest status.

Translation and classification are automated and may differ in nuance from the source.