본문 바로가기

US · 21 CFR

21 CFR 212.71

What actions must I take if a batch of PET drug product does not conform to specifications?

21 CFR Part 212 › Subpart H—Finished Drug Product Controls and Acceptance

조항 원문

(a) Rejection of nonconforming product. You must reject a batch of a PET drug product that does not conform to specifications. You must have and follow procedures to identify and segregate the product to avoid mix-ups. You must have and follow procedures to investigate the cause(s) of the nonconforming product. The investigation must include, but is not limited to, examination of processes, operations, records, complaints, and any other relevant sources of information concerning the nonconforming product.

(b) Investigation. You must document the investigation of a PET drug product that does not meet specifications, including the results of the investigation and what happened to the rejected PET drug product.

(c) Correction of problems. You must take action to correct any identified problems to prevent recurrence of a nonconforming product or other quality problem.

(d) Reprocessing. If appropriate, you may reprocess a batch of a PET drug product that does not conform to specifications. If material that does not meet acceptance criteria is reprocessed, you must follow procedures stated in the product's approved application and the finished product must conform to specifications, except for sterility, before final release.

본문은 원문(영어)입니다. 국문 번역은 준비 중입니다.

출처: U.S. Office of the Federal Register / GPO — eCFR · U.S. federal government work — public domain (17 U.S.C. §105). eCFR is an editorial compilation, not an official legal edition of the CFR. · 수집일 2026-11-16

Weekly Newsletter

FDA·EU·식약처 실사 지적, 매주 받아 보세요

FDA·EMA·식약처의 새 지적과 조치를 공식 원문 링크와 함께 보내드립니다. 사이트에 새로운 기능이 생기면 함께 안내드립니다. 구독은 언제든 한 번의 클릭으로 해지할 수 있습니다.

이번 주 소식 먼저 보기

확인 메일의 버튼을 누르면 구독이 완료됩니다 · 언제든 해지할 수 있습니다