21 CFR 610.42

Restrictions on use for further manufacture of medical devices.

21 CFR Part 610 › Subpart E—Testing Requirements for Relevant Transfusion-Transmitted Infections

조항 원문

(a) In addition to labeling requirements in subchapter H of this chapter, when a medical device contains human blood or a blood component as a component of the final device, and the human blood or blood component was found to be reactive by a screening test performed under § 610.40(a) and (b), then you must include in the device labeling a statement of warning indicating that the product was manufactured from a donation found to be reactive by a screening test for evidence of infection due to the identified relevant transfusion-transmitted infection(s).

(b) FDA may approve an exception or alternative to the statement of warning required in paragraph (a) of this section based on evidence that the reactivity of the human blood or blood component in the medical device presents no significant health risk through use of the medical device.

본문은 원문(영어)입니다. 국문 번역은 준비 중입니다.

각주·출처 표기

  • [66 FR 31164, June 11, 2001, as amended at 80 FR 29897, May 22, 2015]

출처: U.S. Office of the Federal Register / GPO — eCFR · U.S. federal government work — public domain (17 U.S.C. §105). eCFR is an editorial compilation, not an official legal edition of the CFR. · 수집일 2026-11-16