EU GMP Annex 1 4.4

4. Premises

조항 원문

For the manufacture of sterile products, there are four grades of cleanroom/zone.

Grade A: The critical zone for high-risk operations (e.g. aseptic processing line, filling zone, stopper bowl, open primary packaging or for making aseptic connections under the protection of first air). Normally, such conditions are provided by a localised airflow protection, such as unidirectional airflow workstations within RABS or isolators. The maintenance of unidirectional airflow should be demonstrated and qualified across the whole of the grade A area. Direct intervention (e.g. without the protection of barrier and glove port technology) into the grade A area by operators should be minimized by premises, equipment, process and procedural design.

Grade B: For aseptic preparation and filling, this is the background cleanroom for grade A (where it is not an isolator). Air pressure differences should be continuously monitored. Cleanrooms of lower grade than grade B can be considered where isolator technology is used (see paragraph 4.20 ).

Grade C and D: These are cleanrooms used for carrying out less critical stages in the manufacture of aseptically filled sterile products or as a background for isolators. They can also be used for the preparation/filling of terminally sterilised products. (See section 8 for the specific details on terminal sterilisation activities).

본문은 원문(영어)입니다. 국문 번역은 준비 중입니다.

출처: European Commission, DG SANTE · © European Union. Reuse is authorised provided the source is acknowledged (Commission Decision 2011/833/EU on the reuse of Commission documents). · 수집일 2026-11-15

Weekly Newsletter

이런 지적, 매주 월요일 한국어로 받아보세요

FDA·EMA·식약처의 새 지적과 조치를 공식 원문 링크와 함께 보내드립니다. 사이트에 새로운 기능이 생기면 함께 안내드립니다. 구독은 언제든 한 번의 클릭으로 해지할 수 있습니다.

이번 주 소식 먼저 보기

매주 월요일 발송 · 구독 신청 시 확인 메일이 발송됩니다 · 수신거부는 모든 메일 하단에서 가능합니다.