EU GMP Annex 1 8.70
8. Production and Specific Technologies › Dry heat sterilisation
조항 원문
Dry heat ovens are typically employed to sterilise or depyrogenate primary packaging components, starting materials or active substances but may be used for other processes. They should be maintained at a positive pressure relative to lower grade clean areas throughout the sterilisation and post sterilisation hold process unless the integrity of the packaging is maintained. All air entering the oven should pass through a HEPA filter. Critical process parameters that should be considered in qualification and/or routine processing should include, but are not limited to:
i. Temperature.
ii. Exposure period/time.
iii. Chamber pressure (for maintenance of over pressure).
iv. Air speed.
v. Air quality within the oven.
vi. Heat penetration of material/article (slow to heat spots).
vii. Heat distribution/uniformity.
viii. Load pattern and configuration of articles to be sterilised/depyrogenated including minimum and maximum loads.