EU GMP Annex 1 8.82

8. Production and Specific Technologies › Filter sterilisation of products which cannot be sterilised in their final container

조항 원문

The filtration system should be designed to:

i. Allow operation within validated process parameters.

ii. Maintain the sterility of the filtrate.

iii. Minimize the number of aseptic connections required between the final sterilising grade filter and the final filling of the product.

iv. Allow cleaning procedures to be conducted as necessary.

v. Allow sterilisation procedures, including sterilisation in place, to be conducted as necessary.

vi. Permit in-place integrity testing, of the 0.22 µm final sterilising grade filter, preferably as a closed system, both prior to, and following filtration as necessary. In-place integrity testing methods should be selected to avoid any adverse impact on the quality of the product.