EU · EudraLex Vol.4
EU GMP Annex 1 9.49
9. Environmental & process monitoring › Aseptic process simulation (APS) (also known as media fill)
조항 원문
An aseptic process should be subject to a repeat of the initial validation when:
i. The specific aseptic process has not been in operation for an extended period of time.
ii. There is a change to the process, equipment, procedures or environment that has the potential to affect the aseptic process or an addition of new product containers or container-closure combinations.
본문은 원문(영어)입니다. 국문 번역은 준비 중입니다.
출처: European Commission, DG SANTE · © European Union. Reuse is authorised provided the source is acknowledged (Commission Decision 2011/833/EU on the reuse of Commission documents). · 수집일 2026-11-15