EU GMP Annex 1 8.87

8. Production and Specific Technologies › Filter sterilisation of products which cannot be sterilised in their final container

조항 원문

The integrity of the sterilised filter assembly should be verified by integrity testing before use (pre-use post sterilisation integrity test or PUPSIT), to check for damage and loss of integrity caused by the filter preparation prior to use. A sterilising grade filter that is used to sterilise a fluid should be subject to a non-destructive integrity test post-use prior to removal of the filter from its housing. The integrity test process should be validated and test results should correlate to the microbial retention capability of the filter established during validation. Examples of tests that are used include bubble point, diffusive flow, water intrusion or pressure hold test. It is recognized that PUPSIT may not always be possible after sterilisation due to process constraints (e.g. the filtration of very small volumes of solution). In these cases, an alternative approach may be taken providing that a thorough risk assessment has been performed and compliance is achieved by the implementation of appropriate controls to mitigate any risk of a non-integral filtration system. Points to consider in such a risk assessment should include but are not limited to:

i. In depth knowledge and control of the filter sterilisation process to ensure that the potential for damage to the filter is minimized.

ii. In depth knowledge and control of the supply chain to include:

• Contract sterilisation facilities.

• Defined transport mechanisms.

• Packaging of the sterilised filter, to prevent damage to the filter during transportation and storage.

iii. In depth process knowledge such as:

• The specific product type, including particle burden and whether there exists any risk of impact on filter integrity values, such as the potential to alter integrity-testing values and therefore prevent the detection of a non-integral filter during a post-use filter integrity test.

• Pre-filtration and processing steps, prior to the final sterilising grade filter, which would remove particle burden and clarify the product prior to the sterile filtration.

본문은 원문(영어)입니다. 국문 번역은 준비 중입니다.

이 조항을 근거로 든 용어

출처: European Commission, DG SANTE · © European Union. Reuse is authorised provided the source is acknowledged (Commission Decision 2011/833/EU on the reuse of Commission documents). · 수집일 2026-11-15

Weekly Newsletter

이런 지적, 매주 월요일 한국어로 받아보세요

FDA·EMA·식약처의 새 지적과 조치를 공식 원문 링크와 함께 보내드립니다. 사이트에 새로운 기능이 생기면 함께 안내드립니다. 구독은 언제든 한 번의 클릭으로 해지할 수 있습니다.

이번 주 소식 먼저 보기

매주 월요일 발송 · 구독 신청 시 확인 메일이 발송됩니다 · 수신거부는 모든 메일 하단에서 가능합니다.